Working Principle Of Electrosurgery

According to the principle of transistor high-frequency Electrosurgery equipment, after the output components are connected correctly with the control circuit, the working mode and the set output power are selected according to the clinical needs, and the set data is sent to the control circuit.
After connecting the neutral electrode (negative electrode plate), the neutral electrode signal is output by the isolation circuit and sent to the control circuit, allowing the control circuit to operate. The start signal is sent to the control circuit through a isolator through a manual switch or foot switch.
The control circuit outputs corresponding control signals to the modulation circuit, display circuit, high-voltage power supply, alarm circuit, and output drive circuit respectively, making the circuit in working state.
The modulated pulse signal is fed into the isolation amplifier through a logic circuit to modulate the main frequency. The modulated high-frequency signal is transmitted to the power amplifier circuit through the driving pole, and then coupled to the output circuit through an output transformer. At this point, a high-voltage power supply is added to the high-frequency power output pole to provide output energy.
The current limiting detection circuit detects the circuit current and sends sampled data to the control circuit to limit the increase of high-frequency current. Cut off output power when the load is short circuited. Working principle of Electrosurgery